Academic Journal

In Vitro Performance of an Investigational Vibrating-Membrane Nebulizer with Surfactant under Simulated, Non-Invasive Neonatal Ventilation Conditions: Influence of Continuous Positive Airway Pressure Interface and Nebulizer Positioning on the Lung Dose

التفاصيل البيبلوغرافية
العنوان: In Vitro Performance of an Investigational Vibrating-Membrane Nebulizer with Surfactant under Simulated, Non-Invasive Neonatal Ventilation Conditions: Influence of Continuous Positive Airway Pressure Interface and Nebulizer Positioning on the Lung Dose
المؤلفون: Federico Bianco, Elena Pasini, Marcello Nutini, Xabier Murgia, Carolin Stoeckl, Martin Schlun, Uwe Hetzer, Sauro Bonelli, Marta Lombardini, Ilaria Milesi, Marisa Pertile, Stefan Minocchieri, Fabrizio Salomone, Albert Bucholski
المصدر: Pharmaceutics, Vol 12, Iss 3, p 257 (2020)
بيانات النشر: MDPI AG, 2020.
سنة النشر: 2020
المجموعة: LCC:Pharmacy and materia medica
مصطلحات موضوعية: eflow nebulizer, nebulized surfactant, aerosol, non-invasive ventilation, print, Pharmacy and materia medica, RS1-441
الوصف: Non-invasive delivery of nebulized surfactant has been a long-pursued goal in neonatology. Our aim was to evaluate the performance of an investigational vibrating-membrane nebulizer in a realistic non-invasive neonatal ventilation circuit with different configurations. Surfactant (aerosols were generated with a nebulizer in a set-up composed of a continuous positive airway pressure (CPAP) generator with a humidifier, a cast of the upper airway of a preterm infant (PrINT), and a breath simulator with a neonatal breathing pattern. The lung dose (LD), defined as the amount of surfactant collected in a filter placed at the distal end of the PrINT cast, was determined after placing the nebulizer at different locations of the circuit and using either infant nasal mask or nasal prongs as CPAP interfaces. The LD after delivering a range of nominal surfactant doses (100−600 mg/kg) was also investigated. Surfactant aerosol particle size distribution was determined by laser diffraction. Irrespective of the CPAP interface used, about 14% of the nominal dose (200 mg/kg) reached the LD filter. However, placing the nebulizer between the Y-piece and the CPAP interface significantly increased the LD compared with placing it 7 cm before the Y-piece, in the inspiratory limb. (14% ± 2.8 vs. 2.3% ± 0.8, nominal dose of 200 mg/kg). The customized eFlow Neos showed a constant aerosol generation rate and a mass median diameter of 2.7 μm after delivering high surfactant doses (600 mg/kg). The customized eFlow Neos nebulizer showed a constant performance even after nebulizing high doses of undiluted surfactant. Placing the nebulizer between the Y-piece and the CPAP interface achieves the highest LD under non-invasive ventilation conditions.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 1999-4923
Relation: https://www.mdpi.com/1999-4923/12/3/257; https://doaj.org/toc/1999-4923
DOI: 10.3390/pharmaceutics12030257
URL الوصول: https://doaj.org/article/ef6418d36cc64953925b958ec45d4ce2
رقم الانضمام: edsdoj.f6418d36cc64953925b958ec45d4ce2
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:19994923
DOI:10.3390/pharmaceutics12030257