Academic Journal
Misplaced Ventricular Lead via an Atrial Septal Defect–Percutaneous Extraction
العنوان: | Misplaced Ventricular Lead via an Atrial Septal Defect–Percutaneous Extraction |
---|---|
المؤلفون: | Jerry Easo, Malte Book, Samer Hakmi, Alexander Weymann |
المصدر: | The Thoracic & Cardiovascular Surgeon Reports, Vol 08, Iss 01, Pp e8-e10 (2019) |
بيانات النشر: | Georg Thieme Verlag KG, 2019. |
سنة النشر: | 2019 |
المجموعة: | LCC:Surgery |
مصطلحات موضوعية: | lead extraction, misplaced ventricular lead, mitral valve regurgitation, Surgery, RD1-811 |
الوصف: | Abstract Background Necessity for lead removal in this case was after 12 years due to mitral valve regurgitation caused by retraction of the posterior leaflet by the inadvertently misplaced lead. Case Description This history describes a 45-year-old woman with history of multiple cardiac operations at young age with an abandoned defibrillator lead via a patent atrial septal defect. Lead extraction was performed with first described use of rotational dilating sheaths to reduce emboli risk, hoping to avoid a fourth surgical procedure with high risk. Conclusions Percutaneous lead removal using rotational sheaths is possible even for misplaced leads after long-time intervals with acceptable operative risk. |
نوع الوثيقة: | article |
وصف الملف: | electronic resource |
اللغة: | English |
تدمد: | 2194-7635 2194-7643 |
Relation: | https://doaj.org/toc/2194-7635; https://doaj.org/toc/2194-7643 |
DOI: | 10.1055/s-0039-1687821 |
URL الوصول: | https://doaj.org/article/bef2203b332e4e3db25ed9db5ac47615 |
رقم الانضمام: | edsdoj.bef2203b332e4e3db25ed9db5ac47615 |
قاعدة البيانات: | Directory of Open Access Journals |
تدمد: | 21947635 21947643 |
---|---|
DOI: | 10.1055/s-0039-1687821 |