Audit of Anesthetic Equipment in Veterinary Clinics in Spain and Portugal

التفاصيل البيبلوغرافية
العنوان: Audit of Anesthetic Equipment in Veterinary Clinics in Spain and Portugal
المؤلفون: Jose I. Redondo, Luis Domenech, Cristina Mateu, Alfon Bañeres, Amalia Martínez, Diana Lopes
المصدر: Frontiers in Veterinary Science, Vol 7 (2020)
بيانات النشر: Frontiers Media S.A., 2020.
سنة النشر: 2020
المجموعة: LCC:Veterinary medicine
مصطلحات موضوعية: anesthesia, veterinary, equipment malfunction, safety, anesthetic machine, vaporizer, Veterinary medicine, SF600-1100
الوصف: The objective of this retrospective study was to review the results of a 4-year audit performed on anesthetic machines and vaporizers used in veterinary clinics in Spain and Portugal. Data was collected between July 2016 and April 2020. Inspections were carried out by a team of seven veterinarians, using a human-modified system of checks that was adapted to a veterinary practice. The evaluation of each item was noted as “correct” or “incorrect”. The vaporizers' performance was evaluated using a self-calibrating gas analyzer. The vaporizer was classified as “correct” or “incorrect” when the vaporization error was less than or equal to, or more than 20%, respectively. The anesthetic machine was classified as “conforming” if all its components were noted as “correct” and no leaks were detected, or as “non-conforming” if any of the components was noted as “incorrect” or if a leak was detected. If the inspector was able to repair on-site the item malfunctions detected and the machine was fit for use, they issued a final report as “conforming.” On the contrary, if such malfunctions persisted, the final report was “non-conforming,” and a recommendation to remove the machine from service until its final repair was provided. To perform statistical analysis, each inspected item was used as predictor, classification and regression trees were built, and a random forest analysis was performed. A total of 2,001 anesthetic machines and 2,309 vaporizers were studied. After inspection, 42.7 and 26.4% of the machines were non-conforming and conforming, respectively, whereas 30.9% could be repaired in situ. A total of 27.1% of the isoflurane vaporizers and 35.9% of the sevoflurane vaporizers were incorrect. Machine learning techniques showed that the most important variables in the classification of the anesthetic machines as conforming or non-conforming were mostly the scavenger system and the canister, followed some way behind by the APL valve, source of oxygen, reservoir bag, vaporizer, and connections.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2297-1769
Relation: https://www.frontiersin.org/articles/10.3389/fvets.2020.592597/full; https://doaj.org/toc/2297-1769
DOI: 10.3389/fvets.2020.592597
URL الوصول: https://doaj.org/article/80705533c5924f1a9b0b3dfe4a85d387
رقم الانضمام: edsdoj.80705533c5924f1a9b0b3dfe4a85d387
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:22971769
DOI:10.3389/fvets.2020.592597