التفاصيل البيبلوغرافية
العنوان: |
Accuracy Improvement of a Compact 85Rb Atom Gravimeter by Suppressing Laser Crosstalk and Light Shift |
المؤلفون: |
Guiguo Ge, Xi Chen, Jinting Li, Danfang Zhang, Meng He, Wenzhang Wang, Yang Zhou, Jiaqi Zhong, Biao Tang, Jie Fang, Jin Wang, Mingsheng Zhan |
المصدر: |
Sensors, Vol 23, Iss 13, p 6115 (2023) |
بيانات النشر: |
MDPI AG, 2023. |
سنة النشر: |
2023 |
المجموعة: |
LCC:Chemical technology |
مصطلحات موضوعية: |
atom gravimeter, compact, accuracy, stability, Chemical technology, TP1-1185 |
الوصف: |
We design and implement a compact 85Rb atom gravimeter (AG). The diameter of the sensor head is 35 cm and the height is 65 cm; the optical and electronic systems are installed in four standard 3U cabinets. The measurement accuracy of this AG is improved by suppress laser crosstalk and light shift. In addition, the angle of the Raman laser reflector is adjusted and locked, and the attitude of the sensing head is automatically adjusted, and the vibration noise is also compensated. The comparison measurement results between this AG and the superconducting gravimeter indicate that its long-term stability is 0.65 μGal @50000 s. |
نوع الوثيقة: |
article |
وصف الملف: |
electronic resource |
اللغة: |
English |
تدمد: |
1424-8220 |
Relation: |
https://www.mdpi.com/1424-8220/23/13/6115; https://doaj.org/toc/1424-8220 |
DOI: |
10.3390/s23136115 |
URL الوصول: |
https://doaj.org/article/c44780c278fd476db6fb360d8ed0449e |
رقم الانضمام: |
edsdoj.44780c278fd476db6fb360d8ed0449e |
قاعدة البيانات: |
Directory of Open Access Journals |