Academic Journal

Comparison of Two Techniques of Measuring Activated Clotting Time During Cardiac Surgery

التفاصيل البيبلوغرافية
العنوان: Comparison of Two Techniques of Measuring Activated Clotting Time During Cardiac Surgery
المؤلفون: Waqas Alam, Syed Aqeel Hussain, Muhammad Ali, Muhammad Adnan Akram, Muhammad Umar Farooq, Rehan Masroor, Hafsa Khalil, Ayesha Sana
المصدر: Pakistan Armed Forces Medical Journal, Vol 72, Iss SUPPL-3 (2022)
بيانات النشر: Army Medical College Rawalpindi, 2022.
سنة النشر: 2022
المجموعة: LCC:Medicine
LCC:Medicine (General)
مصطلحات موضوعية: Activated clotting time, CABG, Cardiac surgery, Hemochron, Heparin, Medicine, Medicine (General), R5-920
الوصف: Objective: To compare two techniques of activated clotting time (ACT) measurement by two different devices to evaluate the correlation and agreement between these techniques at different points of time during cardiac surgery. Study Design: Comparative Cross-sectional Study. Study settings: Cardiac Anesthesia Department, Armed Forces Institute of Cardiology and National Institute of Heart Diseases, Rawalpindi Pakistan, from Feb 2022 to Apr 2022. Methodology: 100 patients undergoing CABG surgery on pump were included by non-probability consecutive sampling technique. Two devices operating on different techniques of measurement were used to measure ACT at four different points of time during surgery. Same sample was used to measure ACT from both devices. Data was collected and analysed by SPSS V.20.0. Pearson's chi square test and paired sample t-test were applied to measure the significance. Linear regression analysis was done to find correlation and Bland Altman plot was used for checking agreement between ACT values from both devices.p-value less than 0.05 was considered significant. Results: Out of 100, male patients were 79 and female patients were 21. Their mean age was 54.18±11.91 years and mean weight was 79.06±18.49 kg. Moderate positive correlation was found with Pearson r value of 0.55. There was poor agreement between the ACTs measured by two devices. On average Hem R measures ACT 63.54±50.83 seconds higher than Hem JSP. Average mean and median values for Hem R group are 400.03±59.39 sec and 397.75sec respectively and for Hem JSP group are 336.49±44.63 sec and 340.5 sec respectively. Conclusion: There is moderate positive correlation but poor agreement between the values of ACT measured by Hem R and Hem JSP due to different techniques of measurement so these should not be used interchangeably.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 0030-9648
2411-8842
Relation: https://pafmj.org/index.php/PAFMJ/article/view/9550; https://doaj.org/toc/0030-9648; https://doaj.org/toc/2411-8842
DOI: 10.51253/pafmj.v72iSUPPL-3.9550
URL الوصول: https://doaj.org/article/cdaa09a58e4746489caf4b6488c9b6fc
رقم الانضمام: edsdoj.09a58e4746489caf4b6488c9b6fc
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:00309648
24118842
DOI:10.51253/pafmj.v72iSUPPL-3.9550