Academic Journal

Evaluation of setup Uncertainties for Single-Fraction SRS by Comparing Two Different Mask-Creation Methods

التفاصيل البيبلوغرافية
العنوان: Evaluation of setup Uncertainties for Single-Fraction SRS by Comparing Two Different Mask-Creation Methods
المؤلفون: Jong Geun Baek, Hyun Soo Jang, Young Kee Oh, Hyun Jeong Lee, Eng Chan Kim
المساهمون: 오영기, Oh, Young Kee, Dept. of Radiation Oncology (방사선종양학)
بيانات النشر: School of Medicine
سنة النشر: 2015
المجموعة: Keimyung University Medical Library: KUMeL Repository
مصطلحات موضوعية: Setup uncertainties, Systematic uncertainties, BrainLAB mask, Target positioner
الوصف: The purpose of this study was to evaluate the setup uncertainties for single-fraction stereotactic radiosurgery (SF-SRS) based on clinical data with two different mask-creation methods using pretreatment con-beam computed tomography imaging guidance. Dedicated frameless fixation Brain- LAB masks for 23 patients were created as a routine mask (R-mask) making method, as explained in the BrainLAB’s user manual. Alternative masks (A-masks), which were created by modifying the cover range of the R-masks for the patient’s head, were used for 23 patients. The systematic errors including these for each mask and stereotactic target localizer were analyzed, and the errors were calculated as the means ± standard deviations (SD) from the left-right (LR), superior-inferior (SI), anterior-posterior (AP), and yaw setup corrections. In addition, the frequencies of the threedimensional (3D) vector length were analyzed. The values of the mean setup corrections for the R-mask in all directions were < 0.7 mm and < 0.1◦ , whereas the magnitudes of the SDs were relatively large compared to the mean values. In contrast, the means and SDs of the A-mask were smaller than those for the R-mask with the exception of the SD in the AP direction. The means and SDs in the yaw rotational direction for the R-mask and the A-mask system were comparable. 3D vector shifts of larger magnitude occurred more frequently for the R-mask than the A-mask. The setup uncertainties for each mask with the stereotactic localizing system had an asymmetric offset towards the positive AP direction. The A-mask-creation method, which is capable of covering the top of the patient’s head, is superior to that for the R-mask, so the use of the A-mask is encouraged for SF-SRS to reduce the setup uncertainties. Moreover, careful mask-making is required to prevent possible setup uncertainties. ; restriction
نوع الوثيقة: article in journal/newspaper
اللغة: unknown
تدمد: 0374-4884
Relation: Journal of the Korean Physical Society., Vol.67(1) : 38-43, 2015; oak-2015-0116; http://kumel.medlib.dsmc.or.kr/handle/2015.oak/32966
DOI: 10.3938/jkps.67.38
الاتاحة: http://kumel.medlib.dsmc.or.kr/handle/2015.oak/32966
https://doi.org/10.3938/jkps.67.38
Rights: BY_NC_ND ; http://creativecommons.org/licenses/by-nc-nd/2.0/kr
رقم الانضمام: edsbas.EE48110C
قاعدة البيانات: BASE
الوصف
تدمد:03744884
DOI:10.3938/jkps.67.38