Academic Journal

A Comparison of Data Quality and Monitoring Completion Rates Between Clinic and Self-Applied ECG Patches

التفاصيل البيبلوغرافية
العنوان: A Comparison of Data Quality and Monitoring Completion Rates Between Clinic and Self-Applied ECG Patches
المؤلفون: Goergen, Jack A., Peigh, Graham, Hsu, Mike, Wilk, Alan, Nayak, Tanvi, Crosson, Lori, Lenane, Judith, Knight, Bradley P., Passman, Rod
المصدر: Heart Rhythm
بيانات النشر: Published by Elsevier Inc. on behalf of Heart Rhythm Society.
سنة النشر: 2022
مصطلحات موضوعية: Article, info, envir
الوصف: BACKGROUND: Since the onset of the COVID-19 pandemic, direct-to-patient, self-applied ECG patch use has substantially increased. There are limited data comparing clinic versus self-applied ECG patches. OBJECTIVES: To compare rates of ECG patch return, percentages of time patches yielded analyzable data (analyzable time), and percentages of prescribed time ECG patches were worn between clinic and self-applied ECG patches prior to and during COVID-19. METHODS: A retrospective analysis of patients prescribed an ECG patch during “pre-COVID” (3/1/2019-3/1/2020) and “COVID” (4/1/2020-4/1/2021) years was conducted. ECG patch return rates, mean percentages of analyzable time, and mean percentages of prescribed wear time were compared between clinic and self-applied groups. RESULTS: Of 29,093 ECG patch prescriptions (19% COVID self-applied), the COVID self-applied group had a lower return rate (90.8%) than both clinic-applied groups (COVID; 97.1%; pre-COVID; 98.1%; p95% in all groups.
نوع الوثيقة: text
اللغة: English
Relation: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9714183/
الاتاحة: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9714183/
Rights: undefined
رقم الانضمام: edsbas.D5C0BA4F
قاعدة البيانات: BASE