Evaluation of NaI(TL) and plastic scintillators for use in remote, unattended, and portal monitoring

التفاصيل البيبلوغرافية
العنوان: Evaluation of NaI(TL) and plastic scintillators for use in remote, unattended, and portal monitoring
المؤلفون: Staples, P., Audia, J., Bai, Y., Briggs, M., Halbig, J.K., Ianakiev, K.D.
سنة النشر: 2009
المجموعة: SciTec Connect (Office of Scientific and Technical Information - OSTI, U.S. Department of Energy)
مصطلحات موضوعية: 44 INSTRUMENTATION, INCLUDING NUCLEAR AND PARTICLE DETECTORS, 05 NUCLEAR FUELS, SAFEGUARDS, NUCLEAR MATERIALS MANAGEMENT, NAI DETECTORS, PLASTIC SCINTILLATION DETECTORS, COMPARATIVE EVALUATIONS, SENSITIVITY ANALYSIS, REMOTE AREAS, RADIATION MONITORING, EXPERIMENTAL DATA
الوصف: The authors have evaluated and compared some of the relevant operating characteristics of NaI and plastic scintillators for use in various safeguards monitoring applications. These include a sensitivity analysis of the two scintillators to various radiation fields and scintillator response as affected by environmental temperature. A comparison of experiment and modeling via the Monte Carlo N-Particle (MCNP) code has been performed to validate the calculational techniques. This then enables complex detector situations to be simulated with increased confidence.
نوع الوثيقة: other/unknown material
وصف الملف: application/pdf
اللغة: unknown
Relation: http://www.osti.gov/servlets/purl/319715; https://www.osti.gov/biblio/319715; https://doi.org/10.2172/319715
DOI: 10.2172/319715
الاتاحة: http://www.osti.gov/servlets/purl/319715
https://www.osti.gov/biblio/319715
https://doi.org/10.2172/319715
رقم الانضمام: edsbas.A38782E2
قاعدة البيانات: BASE