Academic Journal

Absolute calibration of the Agfa Structurix series films at energies between 2.7 and 6.2 keV

التفاصيل البيبلوغرافية
العنوان: Absolute calibration of the Agfa Structurix series films at energies between 2.7 and 6.2 keV
المؤلفون: Lanier, N. E., Cowan, J. S.
المساهمون: DOE
المصدر: Review of Scientific Instruments ; volume 85, issue 11 ; ISSN 0034-6748 1089-7623
بيانات النشر: AIP Publishing
سنة النشر: 2014
الوصف: Although photo-emulsion technology is many decades old, x-ray film still remains a key asset for diagnosing hydrodynamic features in High-Energy Density (HED) experiments. For decades, the preferred option had been Kodak's direct exposure film. After its discontinuance in 2004, the push to find alternatives began. In many situations, the Agfa Structurix series offers the most favorable substitute, but being new to the HED community, its characterization was lacking. To remedy this, recent experiments, conducted at Brookhaven's National Synchrotron Light Source, provide absolute, monochromatic calibration data for the Agfa Structurix series films at K-shell backlighter energies between 2.7 and 6.2 keV. Absolute response curves are presented for Agfa D8, D7, D4, D4sc, D3, and D2. Moreover, the transmission of each film type is also measured.
نوع الوثيقة: article in journal/newspaper
اللغة: English
DOI: 10.1063/1.4894838
DOI: 10.1063/1.4894838/14069957/11d632_1_online.pdf
الاتاحة: http://dx.doi.org/10.1063/1.4894838
https://pubs.aip.org/aip/rsi/article-pdf/doi/10.1063/1.4894838/14069957/11d632_1_online.pdf
رقم الانضمام: edsbas.8A29346D
قاعدة البيانات: BASE