Wearable impedance monitoring system for dialysis patients

التفاصيل البيبلوغرافية
العنوان: Wearable impedance monitoring system for dialysis patients
المؤلفون: Bonnet, Stephane, Bourgerette, A., Gharbi, S., Rubeck, C., Arkouche, W., Massot, B., Mcadams, Eric, Montalibet, A., Jallon, P.
المساهمون: Commissariat à l'énergie atomique et aux énergies alternatives - Laboratoire d'Electronique et de Technologie de l'Information (CEA-LETI), Direction de Recherche Technologique (CEA) (DRT (CEA)), Commissariat à l'énergie atomique et aux énergies alternatives (CEA)-Commissariat à l'énergie atomique et aux énergies alternatives (CEA), Association pour l'Utilisation du Rein Artificiel Région Lyonnaise (AURAL), INL - Capteurs Biomédicaux (INL - Capteurs), Institut des Nanotechnologies de Lyon (INL), École Centrale de Lyon (ECL), Université de Lyon-Université de Lyon-Université Claude Bernard Lyon 1 (UCBL), Université de Lyon-École Supérieure de Chimie Physique Électronique de Lyon (CPE)-Institut National des Sciences Appliquées de Lyon (INSA Lyon), Université de Lyon-Institut National des Sciences Appliquées (INSA)-Institut National des Sciences Appliquées (INSA)-Centre National de la Recherche Scientifique (CNRS)-École Centrale de Lyon (ECL), Université de Lyon-Institut National des Sciences Appliquées (INSA)-Institut National des Sciences Appliquées (INSA)-Centre National de la Recherche Scientifique (CNRS)
المصدر: 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
https://hal.science/hal-01985483
2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Aug 2016, Orlando, United States. ⟨10.1109/EMBC.2016.7591898⟩
بيانات النشر: HAL CCSD
سنة النشر: 2016
المجموعة: Université de Lyon: HAL
مصطلحات موضوعية: [SPI]Engineering Sciences [physics]
جغرافية الموضوع: Orlando, United States
الوصف: International audience ; This paper describes the development and the validation of a prototype wearable miniaturized impedance monitoring system for remote monitoring in home-based dialysis patients. This device is intended to assess the hydration status of dialysis patients using calf impedance measurements. The system is based on the low-power AD8302 component. The impedance calibration procedure is described together with the Cole parameter estimation and the hydric volume estimation. Results are given on a test cell to validate the design and on preliminary calf measurements showing Cole parameter variations during hemodialysis.
نوع الوثيقة: conference object
اللغة: English
Relation: hal-01985483; https://hal.science/hal-01985483; https://hal.science/hal-01985483/document; https://hal.science/hal-01985483/file/BonnetMontalibet2016.pdf
DOI: 10.1109/EMBC.2016.7591898
الاتاحة: https://hal.science/hal-01985483
https://hal.science/hal-01985483/document
https://hal.science/hal-01985483/file/BonnetMontalibet2016.pdf
https://doi.org/10.1109/EMBC.2016.7591898
Rights: info:eu-repo/semantics/OpenAccess
رقم الانضمام: edsbas.7950206E
قاعدة البيانات: BASE
الوصف
DOI:10.1109/EMBC.2016.7591898