Academic Journal

PREDICTED AND MEASURED OXYGEN CONCENTRATIONS IN THE CIRCLE SYSTEM USING LOW FRESH GAS FLOWS WITH OXYGEN SUPPLIED BY AN OXYGEN CONCENTRATOR

التفاصيل البيبلوغرافية
العنوان: PREDICTED AND MEASURED OXYGEN CONCENTRATIONS IN THE CIRCLE SYSTEM USING LOW FRESH GAS FLOWS WITH OXYGEN SUPPLIED BY AN OXYGEN CONCENTRATOR
المؤلفون: PARKER, C. J. R., SNOWDON, S. L.
بيانات النشر: Oxford University Press
سنة النشر: 1988
المجموعة: HighWire Press (Stanford University)
مصطلحات موضوعية: Original Articles
الوصف: An oxygen concentrator based on a zeolitic molecular sieve has been used to supply piped oxygen in a District General Hospital; such oxygen contains 5% argon. If concentrator oxygen is used in the closed circuit, argon accumulates. The present investigation defines the extent and time course of argon accumulation in partially and fully closed breathing systems. A theoretical model is presented which predicts the extent and rate of inert gas accumulation. The predictions have been tested in a volunteer under laboratory conditions and in five anaesthetized patients. It is recommended that concentrator oxygen may be used for low flow anaesthesia of indefinite duration, provided the fresh oxygen flow to the circuit is at least twice the oxygen consumption. If the circuit is totally closed, periodic opening of the circuit is necessary.
نوع الوثيقة: text
وصف الملف: text/html
اللغة: English
Relation: http://bja.oxfordjournals.org/cgi/content/short/61/4/397; http://dx.doi.org/10.1093/bja/61.4.397
DOI: 10.1093/bja/61.4.397
الاتاحة: http://bja.oxfordjournals.org/cgi/content/short/61/4/397
https://doi.org/10.1093/bja/61.4.397
Rights: Copyright (C) 1988, Oxford University Press
رقم الانضمام: edsbas.757BB02E
قاعدة البيانات: BASE