Calibration of Dissipation Factor Measurements of Capacitance Meters

التفاصيل البيبلوغرافية
العنوان: Calibration of Dissipation Factor Measurements of Capacitance Meters
المؤلفون: Turhan, Enis, Gulmez, Gulay, Erkan, Omer, Hayirli, Cem, Galic, Nesrin, Guler, Nazim A.
سنة النشر: 2018
الوصف: A calibration procedure of absolute dissipation factor calibration for commercial capacitance meters has been developed at TUBITAK UME between dissipation factor values 10(-6) and 10(-1) for the frequencies between 60 Hz and 20 kHz. A guard ring capacitor was employed to determine dissipation factor measurement error of capacitance meter at very low values. Resistors of different values are connected series to capacitor in order to obtain desired dissipation factor measurements, for dissipation factor values higher than 10(-4). In this work, AH2500A and AH2700A ultra precision capacitance bridges were employed as capacitance meters.
نوع الوثيقة: conference object
اللغة: unknown
Relation: https://aperta.ulakbim.gov.tr/record/93189; oai:zenodo.org:93189
الاتاحة: https://aperta.ulakbim.gov.tr/record/93189
Rights: info:eu-repo/semantics/openAccess ; http://www.opendefinition.org/licenses/cc-by
رقم الانضمام: edsbas.5B5F5B48
قاعدة البيانات: BASE