Academic Journal

Testing a new NIF neutron time-of-flight detector with a bibenzyl scintillator on OMEGA

التفاصيل البيبلوغرافية
العنوان: Testing a new NIF neutron time-of-flight detector with a bibenzyl scintillator on OMEGA
المؤلفون: Glebov, V. Yu., Forrest, C., Knauer, J. P., Pruyne, A., Romanofsky, M., Sangster, T. C., Shoup, M. J., Stoeckl, C., Caggiano, J. A., Carman, M. L., Clancy, T. J., Hatarik, R., McNaney, J., Zaitseva, N. P.
المساهمون: U.S. Department of Energy
المصدر: Review of Scientific Instruments ; volume 83, issue 10 ; ISSN 0034-6748 1089-7623
بيانات النشر: AIP Publishing
سنة النشر: 2012
الوصف: A new neutron time-of-flight (nTOF) detector with a bibenzyl crystal as a scintillator has been designed and manufactured for the National Ignition Facility (NIF). This detector will replace a nTOF20-Spec detector with an oxygenated xylene scintillator currently operational on the NIF to improve the areal-density measurements. In addition to areal density, the bibenzyl detector will measure the D–D and D–T neutron yield and the ion temperature of indirect- and direct-drive-implosion experiments. The design of the bibenzyl detector and results of tests on the OMEGA Laser System are presented.
نوع الوثيقة: article in journal/newspaper
اللغة: English
DOI: 10.1063/1.4731001
DOI: 10.1063/1.4731001/14139028/10d309_1_online.pdf
الاتاحة: http://dx.doi.org/10.1063/1.4731001
https://pubs.aip.org/aip/rsi/article-pdf/doi/10.1063/1.4731001/14139028/10d309_1_online.pdf
رقم الانضمام: edsbas.5AF3D4F4
قاعدة البيانات: BASE