Evaluation of a commercial synthetic computed tomography generation solution for magnetic resonance imaging‐only radiotherapy

التفاصيل البيبلوغرافية
العنوان: Evaluation of a commercial synthetic computed tomography generation solution for magnetic resonance imaging‐only radiotherapy
المؤلفون: Gonzalez‐Moya, A., Dufreneix, S., Ouyessad, N., Guillerminet, C., Autret, D.
المصدر: Journal of Applied Clinical Medical Physics ; volume 22, issue 6, page 191-197 ; ISSN 1526-9914 1526-9914
بيانات النشر: Wiley
سنة النشر: 2021
المجموعة: Wiley Online Library (Open Access Articles via Crossref)
الوصف: Purpose To evaluate the Siemens solution generating Synthetic computed tomography (sCT) for magnetic resonance imaging (MRI)‐only radiotherapy (RT). Method A retrospective study was conducted on 47 patients treated with external beam RT for brain or prostate cancer who underwent both MRI and CT for treatment planning. sCT images were generated from MRI using automatic bulk densities segmentation. The geometric accuracy of the sCT was assessed by comparing the Hounsfield Units (HU) difference between sCT and CT for bone structures, soft‐tissue, and full body contour. VMAT plans were computed on the CT for treatment preparation and then copied and recalculated with the same monitor units on the sCT using the AcurosXB algorithm. A 1%‐1mm gamma analysis was performed and DVH metrics for the Planning Target Volume (PTV) like the D mean and the D 98% were compared. In addition, we evaluate the usability of sCT for daily position verification with cone beam computed tomography (CBCT) for 14 prostate patients by comparing sCT/CBCT registration results to CT/CBCT. Results Mean HU differences were small except for the skull (207 HU) and right femoral head of four patients where significant aberrations were found. The mean gamma pass rate was 73.2% for the brain and 84.7% for the prostate and D mean were smaller than 0.5%. Large differences for the D 98% of the prostate group could be correlated to low Dice index of the PTV. The mean difference of translations and rotations were inferior to 3.5 mm and 0.2° in all directions with a major difference in the anterior‐posterior direction. Conclusion The performances of the software were shown to be similar to other sCT generation algorithms in terms of HU difference, dose comparison and daily image localization.
نوع الوثيقة: article in journal/newspaper
اللغة: English
DOI: 10.1002/acm2.13236
الاتاحة: http://dx.doi.org/10.1002/acm2.13236
https://onlinelibrary.wiley.com/doi/pdf/10.1002/acm2.13236
https://onlinelibrary.wiley.com/doi/full-xml/10.1002/acm2.13236
https://aapm.onlinelibrary.wiley.com/doi/pdf/10.1002/acm2.13236
Rights: http://creativecommons.org/licenses/by/4.0/
رقم الانضمام: edsbas.3F2D6FD9
قاعدة البيانات: BASE