Fabrication of hot-wire probes and electronics for constant temperature anemometers
العنوان: | Fabrication of hot-wire probes and electronics for constant temperature anemometers |
---|---|
المؤلفون: | Osorio, Orlando Darío, Silin, Nicolas, Converti, J. |
بيانات النشر: | Planta Piloto de Ingeniería Química |
المجموعة: | CONICET Digital (Consejo Nacional de Investigaciones Científicas y Técnicas) |
مصطلحات موضوعية: | Anemometer Circuit, Hot-Film Sensor, Sensor Fabrication, https://purl.org/becyt/ford/2.2, https://purl.org/becyt/ford/2 |
الوصف: | The purpose of this paper is to discuss the construction of constant temperature hot wire anemometers. Different options are analyzed for the fabrication of sensors and electronics. A practical and simple method is developed to manufacture the probe and associated electronics. The fabricated sensor has a diameter of 2 µm and approximately 500 µm in length. The associated electronics are designed to keep the sensor working at a constant temperature and allow the indirect measurement of speed. The design is simple, inexpensive and highly flexible regarding its use and modifications. The dynamic tests performed and the comparison with a commercial model shows a satisfactory performance. The unit built also includes an option for testing with a square wave, which is generated internally, and a comparator mode to measure the resistance of the sensor. ; Fil: Osorio, Orlando Darío. Comisión Nacional de Energía Atómica. Gerencia del Area de Energía Nuclear. Instituto Balseiro; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina ; Fil: Silin, Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Comisión Nacional de Energía Atómica. Gerencia del Area de Energía Nuclear. Instituto Balseiro; Argentina ; Fil: Converti, J. Comisión Nacional de Energía Atómica. Gerencia del Area de Energía Nuclear. Instituto Balseiro; Argentina |
نوع الوثيقة: | article in journal/newspaper |
وصف الملف: | application/pdf |
اللغة: | English |
تدمد: | 0327-0793 1851-8796 |
Relation: | info:eu-repo/semantics/altIdentifier/url/http://www.laar.uns.edu.ar/indexes/i40_03.htm; info:eu-repo/semantics/altIdentifier/url/http://www.scielo.org.ar/scielo.php?script=sci_arttext&pid=S0327-07932010000300006&lng=es&nrm=iso&tlng=en; http://hdl.handle.net/11336/9488; Osorio, Orlando Darío; Silin, Nicolas; Converti, J.; Fabrication of hot-wire probes and electronics for constant temperature anemometers; Planta Piloto de Ingeniería Química; Latin American Applied Research; 40; 3; 3-2010; 233-239 |
الاتاحة: | http://hdl.handle.net/11336/9488 |
Rights: | info:eu-repo/semantics/openAccess ; https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
رقم الانضمام: | edsbas.3E2835CA |
قاعدة البيانات: | BASE |
تدمد: | 03270793 18518796 |
---|