Academic Journal

Direct Temperature Measurements of Cardiac Stent during MRI Examinations

التفاصيل البيبلوغرافية
العنوان: Direct Temperature Measurements of Cardiac Stent during MRI Examinations
المؤلفون: Acri, G, Campanella, F, Vermiglio, G, Anfuso, C, Testagrossa, B, Cavallaro, D, Urzi, D, Sanzo, A, D'Avanzo, MA, Hartwig, V
المساهمون: Acri, G, Campanella, F, Vermiglio, G, Anfuso, C, Testagrossa, B, Cavallaro, D, Urzi, D, Sanzo, A, D'Avanzo, Ma, Hartwig, V
بيانات النشر: MDPI
سنة النشر: 2023
المجموعة: Università degli Studi di Messina: IRIS
مصطلحات موضوعية: MRI, SAR, cardiac stent, RF deposition, exposure assessment, RF heating
الوصف: Nowadays, Magnetic Resonance Imaging (MRI) is considered the gold standard for imaging the brain, spinal cord, musculoskeletal system, head and neck, and complex congenital heart malformations; consequentially, the number of MRI scans in patients with implantable electronic devices has simultaneously increased. During the entire length of the MRI exam, patients are exposed to electromagnetic fields with different characteristics (static, low frequency, radiofrequency fields), which are related to different risks. The scarce available literature about MRI-induced heating on cardiac stents suggests that excessive temperature rise occurs only in unfavorable cases. Ideally, RF safety assessment could be performed during the anamnestic process, but this simulation process's results are too slow to be performed before patient MRI examination. In this context, we developed a dedicated measurement set-up by focusing our target on the measurement of the heating of a cardiac stent during an MRI examination. Results for the temperature rise trend along the entire stent length during a clinical MRI protocol are shown together with the local Specific Absorption Rate (SAR) values and cumulative equivalent minutes at 43 degrees C (CEM43 degrees C), in order to ensure the safety of patients with MR-conditional devices, also with a view to not inappropriately preclude their access to MRI scans. The obtained results show that the maximum temperature rise (4.12 degrees C) is within the limit of 5 degrees C stated in the stent manual for 15 min of continued scanning with the specific conditions. The maximum temperature rise was in correspondence with the stent tips and calculated SAR confirms the fact that two hotspots are present near the tips of the stent. Finally, the calculated CEM43 degrees C remained well below the proposed threshold for muscle tissue.
نوع الوثيقة: article in journal/newspaper
اللغة: English
Relation: info:eu-repo/semantics/altIdentifier/wos/WOS:001095937200001; volume:13; issue:20; firstpage:1; lastpage:12; numberofpages:12; journal:APPLIED SCIENCES; https://hdl.handle.net/11570/3282968
DOI: 10.3390/app132011414
الاتاحة: https://hdl.handle.net/11570/3282968
https://doi.org/10.3390/app132011414
رقم الانضمام: edsbas.1D7902E1
قاعدة البيانات: BASE