Use of a Pressure Wire for Automatically Correcting Artifacts in Phasic Pressure Tracings From a Fluid-Filled Catheter.

التفاصيل البيبلوغرافية
العنوان: Use of a Pressure Wire for Automatically Correcting Artifacts in Phasic Pressure Tracings From a Fluid-Filled Catheter.
المؤلفون: Johnson, D.T., Svanerud, J., Ahn, J.M., Bezerra, H.G., Collison, D., van 't Veer, M., Hennigan, B., De Bruyne, B., Kirkeeide, R.L., Gould, K.L., Johnson, N.P.
المصدر: Cardiovascular revascularization medicine, vol. 46, pp. 98-105
سنة النشر: 2023
المجموعة: Université de Lausanne (UNIL): Serval - Serveur académique lausannois
مصطلحات موضوعية: Humans, Fractional Flow Reserve, Myocardial, Artifacts, Retrospective Studies, Catheters, Equalization, Fluid-filled catheter, Pressure wire
الوصف: Matching phasic pressure tracings between a fluid-filled catheter and high-fidelity pressure wire has received limited attention, although each part contributes half of the information to clinical decisions. We aimed to study the impact of a novel and automated method for improving the phasic calibration of a fluid-filled catheter by accounting for its oscillatory behavior. Retrospective analysis of drift check tracings was performed using our algorithm that corrects for mean difference (offset), temporal delays (timing), differential sensitivity of the manifold transducer and pressure wire sensor (gain), and the oscillatory behavior of the fluid-filled catheter described by its resonant frequency and damping factor (how quickly oscillations disappear after a change in pressure). Among 2886 cases, correcting for oscillations showed a large improvement in 28 % and a medium improvement in 41 % (decrease in root mean square error >0.5 mmHg to <1 or 1-2 mmHg, respectively). 96 % of oscillators were underdamped with median damping factor 0.27 and frequency 10.6 Hz. Fractional flow reserve or baseline Pd/Pa demonstrated no clinically important bias when ignoring oscillations. However, uncorrected subcycle non-hyperemic pressure ratios (NHPR) displayed both bias and scatter. By automatically accounting for the oscillatory behavior of a fluid-filled catheter system, phasic matching against a high-fidelity pressure wire can be improved compared to standard equalization methods. The majority of tracings contain artifacts, mainly due to underdamped oscillations, and neglecting them leads to biased estimates of equalization parameters. No clinically important bias exists for whole-cycle metrics, in contrast to significant effects on subcycle NHPR.
نوع الوثيقة: article in journal/newspaper
وصف الملف: application/pdf
اللغة: English
Relation: info:eu-repo/semantics/altIdentifier/pmid/35918253; info:eu-repo/semantics/altIdentifier/eissn/1878-0938; info:eu-repo/semantics/altIdentifier/urn/urn:nbn:ch:serval-BIB_E2D7ABC5B3BF1; https://serval.unil.ch/notice/serval:BIB_E2D7ABC5B3BF; https://serval.unil.ch/resource/serval:BIB_E2D7ABC5B3BF.P001/REF.pdf
DOI: 10.1016/j.carrev.2022.07.021
الاتاحة: https://serval.unil.ch/notice/serval:BIB_E2D7ABC5B3BF
https://doi.org/10.1016/j.carrev.2022.07.021
https://serval.unil.ch/resource/serval:BIB_E2D7ABC5B3BF.P001/REF.pdf
http://nbn-resolving.org/urn/resolver.pl?urn=urn:nbn:ch:serval-BIB_E2D7ABC5B3BF1
Rights: info:eu-repo/semantics/openAccess ; CC BY 4.0 ; https://creativecommons.org/licenses/by/4.0/
رقم الانضمام: edsbas.18280384
قاعدة البيانات: BASE