[Research on the method of interference correction for nondispersive infrared multi-component gas analysis]

التفاصيل البيبلوغرافية
العنوان: [Research on the method of interference correction for nondispersive infrared multi-component gas analysis]
المؤلفون: You-Wen, Sun, Wen-Qing, Liu, Shi-Mei, Wang, Shu-Hua, Huang, Xiao-Man, Yu
المصدر: Guang pu xue yu guang pu fen xi = Guang pu. 31(10)
سنة النشر: 2012
الوصف: A method of interference correction for nondispersive infrared multi-component gas analysis was described. According to the successive integral gas absorption models and methods, the influence of temperature and air pressure on the integral line strengths and linetype was considered, and based on Lorentz detuning linetypes, the absorption cross sections and response coefficients of H2O, CO2, CO, and NO on each filter channel were obtained. The four dimension linear regression equations for interference correction were established by response coefficients, the absorption cross interference was corrected by solving the multi-dimensional linear regression equations, and after interference correction, the pure absorbance signal on each filter channel was only controlled by the corresponding target gas concentration. When the sample cell was filled with gas mixture with a certain concentration proportion of CO, NO and CO2, the pure absorbance after interference correction was used for concentration inversion, the inversion concentration error for CO2 is 2.0%, the inversion concentration error for CO is 1.6%, and the inversion concentration error for NO is 1.7%. Both the theory and experiment prove that the interference correction method proposed for NDIR multi-component gas analysis is feasible.
تدمد: 1000-0593
URL الوصول: https://explore.openaire.eu/search/publication?articleId=pmid________::e5aae21ddb510eef5f7449fbe4797587
https://pubmed.ncbi.nlm.nih.gov/22250543
رقم الانضمام: edsair.pmid..........e5aae21ddb510eef5f7449fbe4797587
قاعدة البيانات: OpenAIRE