Results of a prospective multicenter study for evaluation of the diagnostic quality of an open whole-body low-field MRI unit. A comparison with high-field MRI measured by the applicable gold standard

التفاصيل البيبلوغرافية
العنوان: Results of a prospective multicenter study for evaluation of the diagnostic quality of an open whole-body low-field MRI unit. A comparison with high-field MRI measured by the applicable gold standard
المؤلفون: T, Merl, M, Scholz, P, Gerhardt, M, Langer, J, Laubenberger, H D, Weiss, H B, Gehl, K J, Wolf, I, Ohnesorge
المصدر: European journal of radiology. 30(1)
سنة النشر: 1999
مصطلحات موضوعية: Adult, Aged, 80 and over, Male, Brain Diseases, Shoulder, Middle Aged, Magnetic Resonance Imaging, Kidney Neoplasms, ROC Curve, Evaluation Studies as Topic, Humans, Female, Spinal Diseases, Prospective Studies, Joint Diseases, Aged
الوصف: To evaluate the diagnostic quality of an open whole-body low-field MRI scanner compared to high-field scanners.Over a period of 3 months, 401 patients with diseases of the kidney (n = 78), the shoulder (n = 122), the spine (n = 105) and the cerebrum (n = 96) were prospectively evaluated in four participating centers. They all underwent clinical evaluation, low-field and high-field MRI examination and surgical or follow-up confirmation of diagnosis. Clinical, histopathologic, high-field and low-field MRI diagnoses were recorded in standardized questionnaires that were centrally evaluated. Statistical evaluation comprised two parts: ROC analysis assessed accuracy of MRI and clinical diagnoses; furthermore rates of concordance of high- and low-field MRI diagnosis were calculated.We found no statistically relevant difference in high-field MRI diagnosis compared to low-field MRI diagnostic accuracy measured by clinical or surgical gold standard in three of the four regions examined; in cerebral examinations there was a small yet significant advantage for the high-field systems (P = 0.01).We conclude that the open low-field scanner we evaluated using clinical and surgical gold standard as reference is able to achieve comparable diagnostic accuracy compared to high-field scanners at lower costs and greater patient comfort. Limitations due to field strength (signal-to-noise ratio, resolution, scan time) seem to be relevant only in a very small number of cases that warrant high-field examination.
تدمد: 0720-048X
URL الوصول: https://explore.openaire.eu/search/publication?articleId=pmid________::97f6ec7c848bc1ab95b18a19bee185d7
https://pubmed.ncbi.nlm.nih.gov/10389012
رقم الانضمام: edsair.pmid..........97f6ec7c848bc1ab95b18a19bee185d7
قاعدة البيانات: OpenAIRE