Determining the optimal paddle force for external defibrillation

التفاصيل البيبلوغرافية
العنوان: Determining the optimal paddle force for external defibrillation
المؤلفون: Graham W. Petley, Daniel M Sado, Frank Clewlow, Charles D. Deakin
المصدر: The American journal of cardiology. 90(7)
سنة النشر: 2002
مصطلحات موضوعية: medicine.medical_specialty, business.industry, Defibrillation, medicine.medical_treatment, Electric Countershock, Myocardial Infarction, Respiration, Artificial, Control theory, Internal medicine, Anesthesia, Practice Guidelines as Topic, Cardiology, Pressure, Paddle, Medicine, Humans, Cardiology and Cardiovascular Medicine, business
الوصف: In summary, 8-kg paddle force achieves 95% of the overall attainable decrease in TTI when using adult defibrillation paddles placed in an anterior-apical position. We recommend that this is the minimum force that should be applied, with a maximum force of 12 kg if physically possible.