التفاصيل البيبلوغرافية
العنوان:
Determining the optimal paddle force for external defibrillation
المؤلفون:
Graham W. Petley , Daniel M Sado , Frank Clewlow , Charles D. Deakin
المصدر:
The American journal of cardiology . 90(7)
سنة النشر:
2002
مصطلحات موضوعية:
medicine.medical_specialty , business.industry , Defibrillation , medicine.medical_treatment , Electric Countershock , Myocardial Infarction , Respiration, Artificial , Control theory , Internal medicine , Anesthesia , Practice Guidelines as Topic , Cardiology , Pressure , Paddle , Medicine , Humans , Cardiology and Cardiovascular Medicine , business
الوصف:
In summary, 8-kg paddle force achieves 95% of the overall attainable decrease in TTI when using adult defibrillation paddles placed in an anterior-apical position. We recommend that this is the minimum force that should be applied, with a maximum force of 12 kg if physically possible.