Time-correlated Single Ion Counting Mass Spectrometer with Long and Short Time-of-Flight Tubes and an Evaluation of Its Performance for Use in Trace Analysis of Allergenic Substances
العنوان: | Time-correlated Single Ion Counting Mass Spectrometer with Long and Short Time-of-Flight Tubes and an Evaluation of Its Performance for Use in Trace Analysis of Allergenic Substances |
---|---|
المؤلفون: | Tomoko Imasaka, Totaro Imasaka, Tiantian Ju, Hiroshi Nakamura, Katsunori Yoshinaga |
المصدر: | Analytical Sciences. 36:539-543 |
بيانات النشر: | Springer Science and Business Media LLC, 2020. |
سنة النشر: | 2020 |
مصطلحات موضوعية: | Laser spectrometry, Chemistry, 010401 analytical chemistry, Analytical chemistry, 02 engineering and technology, 021001 nanoscience & nanotechnology, Laser, Mass spectrometry, 01 natural sciences, Signal, 0104 chemical sciences, Analytical Chemistry, law.invention, Time of flight, Analog signal, law, Ionization, Femtosecond, 0210 nano-technology |
الوصف: | An analyte molecule was ionized using a femtosecond laser as the ionization source and was measured by a twin-type time-of-flight mass spectrometer with long (42 cm) and short (6.4 cm) flight tubes. The signal was measured using an analog signal digitizer and a time-correlated single ion counting system, and performance was evaluated by comparing data obtained from both instruments. The short mass spectrometer had a mass resolution of 450 and was used in the trace analysis of allergenic substances in a fragrance. |
تدمد: | 1348-2246 0910-6340 |
DOI: | 10.2116/analsci.19sbp03 |
URL الوصول: | https://explore.openaire.eu/search/publication?articleId=doi_dedup___::7f72a204258694f5e5d079ea56c1befa https://doi.org/10.2116/analsci.19sbp03 |
Rights: | OPEN |
رقم الانضمام: | edsair.doi.dedup.....7f72a204258694f5e5d079ea56c1befa |
قاعدة البيانات: | OpenAIRE |
تدمد: | 13482246 09106340 |
---|---|
DOI: | 10.2116/analsci.19sbp03 |