Time-correlated Single Ion Counting Mass Spectrometer with Long and Short Time-of-Flight Tubes and an Evaluation of Its Performance for Use in Trace Analysis of Allergenic Substances

التفاصيل البيبلوغرافية
العنوان: Time-correlated Single Ion Counting Mass Spectrometer with Long and Short Time-of-Flight Tubes and an Evaluation of Its Performance for Use in Trace Analysis of Allergenic Substances
المؤلفون: Tomoko Imasaka, Totaro Imasaka, Tiantian Ju, Hiroshi Nakamura, Katsunori Yoshinaga
المصدر: Analytical Sciences. 36:539-543
بيانات النشر: Springer Science and Business Media LLC, 2020.
سنة النشر: 2020
مصطلحات موضوعية: Laser spectrometry, Chemistry, 010401 analytical chemistry, Analytical chemistry, 02 engineering and technology, 021001 nanoscience & nanotechnology, Laser, Mass spectrometry, 01 natural sciences, Signal, 0104 chemical sciences, Analytical Chemistry, law.invention, Time of flight, Analog signal, law, Ionization, Femtosecond, 0210 nano-technology
الوصف: An analyte molecule was ionized using a femtosecond laser as the ionization source and was measured by a twin-type time-of-flight mass spectrometer with long (42 cm) and short (6.4 cm) flight tubes. The signal was measured using an analog signal digitizer and a time-correlated single ion counting system, and performance was evaluated by comparing data obtained from both instruments. The short mass spectrometer had a mass resolution of 450 and was used in the trace analysis of allergenic substances in a fragrance.
تدمد: 1348-2246
0910-6340
DOI: 10.2116/analsci.19sbp03
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::7f72a204258694f5e5d079ea56c1befa
https://doi.org/10.2116/analsci.19sbp03
Rights: OPEN
رقم الانضمام: edsair.doi.dedup.....7f72a204258694f5e5d079ea56c1befa