Quick, easy, cheap, effective, rugged and safe method with magnetic graphitized carbon black and primary secondary amine as adsorbent and its application in pesticide residue analysis

التفاصيل البيبلوغرافية
العنوان: Quick, easy, cheap, effective, rugged and safe method with magnetic graphitized carbon black and primary secondary amine as adsorbent and its application in pesticide residue analysis
المؤلفون: Hao-Bo Zheng, Yun-Qing Huang, Qiong-Wei Yu, Jie-Zhen Mo, Qin Zhao, Luo Yanbo, Yu-Qi Feng
المصدر: Journal of chromatography. A. 1300
سنة النشر: 2013
مصطلحات موضوعية: Detection limit, Analyte, Chromatography, Pesticide residue, Chemistry, Organic Chemistry, Pesticide Residues, Reproducibility of Results, General Medicine, Carbon black, Quechers, Biochemistry, Gas Chromatography-Mass Spectrometry, Analytical Chemistry, Adsorption, Soot, Limit of Detection, Vegetables, Linear Models, Gas chromatography, Gas chromatography–mass spectrometry, Amines, Magnetite Nanoparticles
الوصف: By using magnetic graphitized carbon black and primary secondary amine (GCB/PSA/MNPs) as adsorbent, a modified quick, easy, cheap, effective, rugged and safe (QuEChERS) method was proposed for pesticide residue analysis in vegetables. The magnetic adsorbent was fabricated via simple co-mixing method based on an “aggregate warp” mechanism. To achieve the optimum conditions of modified QuEChERS toward target analytes, several parameters, including the composition of analyte protectants and the amount of the adsorbents were investigated. Under the optimized conditions, a simple, rapid and effective method for the determination of 10 pesticide residues in vegetables was established by coupling modified QuEChERS to gas chromatography/mass spectrometry analysis. The detection limits of the proposed method for 10 pesticides ranged from 0.39 to 8.6 ng/g. Good linearity ( R value ≥ 0.990) was achieved at concentration levels of 10–200 ng/g, and acceptable method reproducibility was found as intra- and inter-day precisions, yielding the relative standard deviations less than 10.7% and 13.4%, respectively. The recoveries were in the range of 69.9–125.0% at different concentrations for real samples. Compared with the reported methods for the determination of a large number of samples, the proposed method has the advantage of less time-consuming in clean-up procedure.
تدمد: 1873-3778
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::78557455091fb7bb1847973940e0ec80
https://pubmed.ncbi.nlm.nih.gov/23659978
رقم الانضمام: edsair.doi.dedup.....78557455091fb7bb1847973940e0ec80
قاعدة البيانات: OpenAIRE