A New Coupling Method for Accurate Measurement of Pedicle Screw Electrical Properties for Surgical Procedures
العنوان: | A New Coupling Method for Accurate Measurement of Pedicle Screw Electrical Properties for Surgical Procedures |
---|---|
المؤلفون: | Pedro Fonseca, João Paulo Vilas-Boas, Miguel V. Correia, Manuel Gutierres, Márcio Fagundes Goethel, Ricardo Sebastião, Manoela Vieira Sousa |
المصدر: | Applied Sciences Volume 11 Issue 21 Applied Sciences, Vol 11, Iss 9861, p 9861 (2021) |
بيانات النشر: | Multidisciplinary Digital Publishing Institute, 2021. |
سنة النشر: | 2021 |
مصطلحات موضوعية: | Technology, Materials science, QH301-705.5, QC1-999, intraoperative monitoring, spine surgery, Electrical resistance and conductance, Electrical resistivity and conductivity, Electrical analysis, Coupling (piping), General Materials Science, Biology (General), Pedicle screw, electrical stimulation, Instrumentation, QD1-999, Fluid Flow and Transfer Processes, Process Chemistry and Technology, Physics, General Engineering, pedicle instrumentation, Titanium alloy, Surgical procedures, pedicle screw, Engineering (General). Civil engineering (General), Computer Science Applications, Chemistry, electrical resistance, TA1-2040, Multimeter, Biomedical engineering |
الوصف: | The objective of this study is to present a new coupling method in order to measure the electrical properties of titanium alloy pedicle screws used in spinal surgery and to compare it with other common methods of measurement. An experimental setup was devised to test the electrical resistance of two specimens of pedicle screws using four methods for coupling the sensing leads, including the use of multimeter probes, alligator clips, wrapped wires and encapsulation with thermo-retractable sleeves. The electrical resistance of the pedicle screw under testing was measured at a current of 10 mA for each coupling method, and the results compared. Our findings show that although widely used in electrical analysis, the alligator clips do not perform as well as the other methods, such as simple wrapping of wires around the screw or the direct application of multimeter probes. The use of thermo-retractable sleeves provides the lowest resistance and inter-quartile range and is closer to the tabled values for the screw’s titanium alloy. Additionally, only this method allows the measurement of identical resistivity values between different screw models manufactured with the same titanium alloy. We then concluded that the use of wrapped wires encapsulated with thermo-retractable sleeves allow more accurate measurements of the pedicle screw’s electrical properties. |
وصف الملف: | application/pdf |
اللغة: | English |
تدمد: | 2076-3417 |
DOI: | 10.3390/app11219861 |
URL الوصول: | https://explore.openaire.eu/search/publication?articleId=doi_dedup___::4ab67194a50280d218847767a818744a |
Rights: | OPEN |
رقم الانضمام: | edsair.doi.dedup.....4ab67194a50280d218847767a818744a |
قاعدة البيانات: | OpenAIRE |
تدمد: | 20763417 |
---|---|
DOI: | 10.3390/app11219861 |