Resonance frequency measurementsin vivoand related surface properties of magnesium-incorporated, micropatterned and magnesium-incorporated TiUnite®, Osseotite®, SLA®and TiOblast®implants

التفاصيل البيبلوغرافية
العنوان: Resonance frequency measurementsin vivoand related surface properties of magnesium-incorporated, micropatterned and magnesium-incorporated TiUnite®, Osseotite®, SLA®and TiOblast®implants
المؤلفون: Carina B. Johansson, Goui-Seong Yoon, Young-Taeg Sul, Jörgen Jönsson
المصدر: Clinical Oral Implants Research. 20:1146-1155
بيانات النشر: Wiley, 2009.
سنة النشر: 2009
مصطلحات موضوعية: Dental Stress Analysis, medicine.medical_specialty, Materials science, Implant surface, Surface Properties, chemistry.chemical_element, Vibration, Osseointegration, Dental Prosthesis Retention, Implants, Experimental, Bone Density, In vivo, medicine, Animals, Magnesium, Dental Restoration Failure, Dental Implants, Titanium, Tibia, Dental Implantation, Endosseous, Implant stability quotient, Surgery, Resonance frequency analysis, Dental Prosthesis Design, chemistry, Rabbits, Stress, Mechanical, Implant, Oral Surgery, Dental Alloys, Biomedical engineering
الوصف: OBJECTIVE: To investigate implant stability using resonance frequency measurements of topographically changed and/or surface chemistry-modified implants in rabbit bone. MATERIAL AND METHODS: Six groups of microstructured, screw-shaped titanium implants: two oxidized, cation-incorporated experimental implants [Mg implants and MgMp implants with micropatterned thread flanges (80-150 microm wide and 60-70 microm deep)] and four commercially available clinical implants (TiUnite((R)), Osseotite((R)), SLA((R)), and TiOblast((R))) were installed in 10 rabbit tibia for 6 weeks. The surface properties of the implants were characterized in detail using several analytical techniques. Implant stability was measured using a resonance frequency analyzer (Osstell(TM)). RESULTS: Surface characterization of the implants revealed microstructured, moderately rough implant surfaces varying 0.7-1.4 mum in S(a) (mean height deviation), but with clear differences in surface chemistry. After 6 weeks, all implants showed statistically significantly higher increases in implant stability. When compared with one another, MgMp implants showed the most significant mean implant stability quotient (ISQ) value relative to the others (P
تدمد: 1600-0501
0905-7161
DOI: 10.1111/j.1600-0501.2009.01767.x
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::1495d4e5593235cb93e6875be50d925f
https://doi.org/10.1111/j.1600-0501.2009.01767.x
Rights: CLOSED
رقم الانضمام: edsair.doi.dedup.....1495d4e5593235cb93e6875be50d925f
قاعدة البيانات: OpenAIRE
الوصف
تدمد:16000501
09057161
DOI:10.1111/j.1600-0501.2009.01767.x