Performance of Multisphere Spectrometry Systems
العنوان: | Performance of Multisphere Spectrometry Systems |
---|---|
المؤلفون: | Jeremy Ed Sweezy, K. G. Veinot, Nolan E. Hertel, R.A. Karam |
المصدر: | Radiation Protection Dosimetry. 78:263-272 |
بيانات النشر: | Oxford University Press (OUP), 1998. |
سنة النشر: | 1998 |
مصطلحات موضوعية: | Bonner sphere, Radiation, Materials science, Radiological and Ultrasound Technology, Spectrometer, Physics::Instrumentation and Detectors, business.industry, Astrophysics::High Energy Astrophysical Phenomena, Nuclear engineering, Physics::Medical Physics, Public Health, Environmental and Occupational Health, General Medicine, Scintillator, Neutron flux, Calibration, Radiology, Nuclear Medicine and imaging, Neutron, Research reactor, Thermoluminescent dosimeter, Nuclear Experiment, Nuclear medicine, business |
الوصف: | A TLD-based multisphere spectrometer has been constructed to measure neutron spectra under high fluence-rate conditions. The multisphere system uses paircd TLD-600 and TLD-700 thermoluminescence detectors. Spectral measurements made with this system are compared with measurements made with a 6 Lil(Eu) scintillator-based Bonner sphere spectrometer. The system was used to measure the neutron spectrum of the Georgia Tech Research Reactor Bio-Medical Facility under high neutron fluence rates. The scintillator-based system was used at low neutron and gamma fluence rates. A comparison of this unfolded spcctrum is made with an MCNP codc calculation performed using a simplified representation of the reactor and the bcam facility. |
تدمد: | 1742-3406 0144-8420 |
DOI: | 10.1093/oxfordjournals.rpd.a032359 |
URL الوصول: | https://explore.openaire.eu/search/publication?articleId=doi_________::88d2bc5d0bb2e5a5bdc72381ecda9498 https://doi.org/10.1093/oxfordjournals.rpd.a032359 |
رقم الانضمام: | edsair.doi...........88d2bc5d0bb2e5a5bdc72381ecda9498 |
قاعدة البيانات: | OpenAIRE |
تدمد: | 17423406 01448420 |
---|---|
DOI: | 10.1093/oxfordjournals.rpd.a032359 |