مستخلص: |
In this study, we thoroughly analyzed the linearity and repeatability of force-sensing resistor (FSR) sensors through static load tests to ensure their reliability. The novelty of this research lies in its comprehensive evaluation and direct comparison of two widely used FSR sensors, i.e., Flexiforce A201-1 and Interlink FSR-402, under various loading conditions by employing a robust calibration methodology. This study provides detailed insights into the sensors' performances, offering practical calibration equations that enhance measurement precision and reliability, which have not been extensively documented in previous studies. Our results demonstrate that the linearity of thin film FSR sensors is highly accurate, closely resembling a straight line. We employed M1 Class weights, applying loads ranging from 20 g to 300 g. The resistance of the FSR sensors, which varies with the applied load, was measured using a voltage divider circuit and an analog-to-digital converter of a microcontroller. MATLAB was used to calculate the average output voltage for each applied load and fixed resistance. Additionally, we examined the relationships among load, FSR sensor resistance, and conductivity. Our research indicates that with precise calibration, thin film FSR sensors can be highly reliable for force measurement applications. [ABSTRACT FROM AUTHOR] |
|
Copyright of Applied Sciences (2076-3417) is the property of MDPI and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
قاعدة البيانات: |
Complementary Index |