基于EGSnrc程序的锥形束CT模拟仿真研究. (Chinese) |
Alternate Title: |
Cone beam CT simulation research based on EGSnrc program. (English) |
المؤلفون: |
付娟, 孟慧鹏, 艾克文, 顾欣, 王克强, 张艳龙, 丁红军, 段敬豪, 孙倩 |
المصدر: |
Chinese Medical Equipment Journal; 2019, Vol. 40 Issue 10, p29-52, 6p |
Abstract (English): |
Objective To execute cone beam CT simulation based on EGSnrc program and verify the accuracy of the developed system. Methods Cone beam CT was divided into four modules of X-ray source, ray source component, detector and Monte Carlo simulation phantom with BEAMnre subroutine of EGSnrc program, and then modeling was carried out accordingly. The results of X-ray energy spectrum, percentage depth dose and off-axis dose by the system developed were analyzed and compared with those by the third-party software and actual measurement to verify the accuracy of the system. Results There was a high consistency between the spectrum of the system and that by Speklr program, the maximum difference was 3.0% and the mean difference was lower than 2.0% ; Monte Carlo simulation had the results of percentage depth dose as well as transverse and longitudinal off-axis dose with high consistencies with those by actual measurement, the maximum differences were 2.7%. 4.0% and 2.7% respectively and the mean differences were restrained within ±2.0%, ±1.1% and ±0.8% respectively. Conclusion The established simulation system realizes accurate simulation of cone beam CT, and lays a foundation for the researches on cone beam CT. [ABSTRACT FROM AUTHOR] |
Abstract (Chinese): |
目的:基于EGSnrc程序实现对锥形束CT系统的模拟仿真,并对所建立系统的准确性进行验证。方法 :通过EGSnrc程序的BEAMnrc子程序将锥形束CT系统分为X射线源、射线源组件、探测器、Monte Carlo模拟用模体4个模块并对其进行建模。对所建立的模拟仿真系统的X射线能谱、百分深度剂量、离轴剂量与第三方软件或实测结果进行分析比较,以验证其准确性。结果:模拟仿真系统的能谱与Spektr程序生成的能谱有良好的一致性,最大差异为3.0%,平均差异小于2.0%;Monte Carlo模拟的百分深度剂量和横、纵向离轴剂量结果与实测结果的一致性良好,最大差异分别为2.7%、4.0%、2.7%,平均差异分别在±2.0%、±1.1%、±0.8%以内。结论:建立的模拟仿真系统可实现锥形束CT的准确模拟,可为研究锥形束CT的相关问题奠定基础。 [ABSTRACT FROM AUTHOR] |
|
Copyright of Chinese Medical Equipment Journal is the property of Chinese Medical Equipment Journal Editorial Office and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
قاعدة البيانات: |
Complementary Index |