Elecsys® and Kryptor immunoassays for the measurement of sFlt-1 and PlGF to aid preeclampsia diagnosis: are they comparable?

التفاصيل البيبلوغرافية
العنوان: Elecsys® and Kryptor immunoassays for the measurement of sFlt-1 and PlGF to aid preeclampsia diagnosis: are they comparable?
المؤلفون: Stepan, Holger, Hund, Martin, Dilba, Peter, Sillman, Johanna, Schlembach, Dietmar
المصدر: Clinical Chemistry & Laboratory Medicine; Sep2019, Vol. 57 Issue 9, p1339-1348, 10p, 2 Charts, 3 Graphs
مصطلحات موضوعية: PREECLAMPSIA, IMMUNOASSAY
مستخلص: Background: For pregnant women with suspected preeclampsia, the soluble fms-like tyrosine-kinase 1 (sFlt-1)/placental growth factor (PlGF) ratio is a biomarker to aid diagnosis. We performed method comparisons between Elecsys® and Kryptor sFlt-1 and PlGF immunoassays and assessed the diagnostic performance for preeclampsia. Methods: Serum samples from a case-control study involving 113 pregnant women with preeclampsia/elevated liver enzymes and low platelet count (HELLP) and 270 controls were analyzed. sFlt-1 and PlGF were measured using Roche Elecsys® and BRAHMS Kryptor sFlt-1/PlGF immunoassays. The sFlt-1/PlGF ratios were calculated, and Passing-Bablok regression/Bland-Altman plots were performed. Gestation-specific cut-offs, ≤33 and ≥85/≥110, were assessed. Results: Mean (±2 standard deviation [SD]) differences between the Elecsys® and Kryptor values were: sFlt-1, 173.13 pg/mL (6237.66, −5891.40); PlGF, −102.71 pg/mL (186.06, −391.48); and sFlt-1/PlGF, 151.74 (1085.11, −781.63). The Elecsys® and Kryptor immunoassays showed high correlation: Pearson's correlation coefficients were 0.913 (sFlt-1) and 0.945 (PlGF). Slopes were 1.06 (sFlt-1) and 0.79 (PlGF), resulting in ~20% lower values for Kryptor PlGF. Sensitivities and specificities using the sFlt-1/PlGF ≥85 cut-off for early-onset preeclampsia (20 + 0 to 33 + 6 weeks) were 88.1%/100.0% (Elecsys®) and 90.5%/96.2% (Kryptor), respectively, and using the ≥110 cut-off for late-onset preeclampsia (≥34 + 0 weeks) were 51.3%/96.5% (Elecsys®) and 78.9%/90.1% (Kryptor), respectively. Using Elecsys® and Kryptor sFlt-1/PlGF, 0% and 3.8% of women, respectively, were falsely ruled-in for early-onset, and 3.5% and 9.9%, respectively, for late-onset preeclampsia. Conclusions: Despite high correlation between the Elecsys® and Kryptor immunoassays, we observed significant differences between sFlt-1/PlGF and PlGF results. Therefore, sFlt-1/PlGF cut-offs validated for Elecsys® immunoassays are not transferable to Kryptor immunoassays. [ABSTRACT FROM AUTHOR]
Copyright of Clinical Chemistry & Laboratory Medicine is the property of De Gruyter and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:14346621
DOI:10.1515/cclm-2018-1228