التفاصيل البيبلوغرافية
العنوان: |
Five‐analyzer Johann spectrometer for hard X‐ray photon‐in/photon‐out spectroscopy at the Inner Shell Spectroscopy beamline at NSLS‐II: design, alignment and data acquisition. |
المؤلفون: |
Tayal, Akhil1 (AUTHOR), Coburn, David Scott1 (AUTHOR), Abel, Donald1 (AUTHOR), Rakitin, Max1 (AUTHOR), Ivashkevych, Oksana1 (AUTHOR), Wlodek, Jakub1 (AUTHOR), Wierzbicki, Dominik1,2 (AUTHOR), Xu, Weihe1 (AUTHOR), Nazaretski, Evgeny1 (AUTHOR), Stavitski, Eli1 (AUTHOR), Leshchev, Denis1 (AUTHOR) leshchev.denis@gmail.com |
المصدر: |
Journal of Synchrotron Radiation. Nov2024, Vol. 31 Issue 6, p1609-1621. 13p. |
مصطلحات موضوعية: |
*MATERIALS science, *LIGHT sources, *ELECTRONIC structure, *ACQUISITION of data, *ENVIRONMENTAL sciences |
مستخلص: |
Here, a recently commissioned five‐analyzer Johann spectrometer at the Inner Shell Spectroscopy beamline (8‐ID) at the National Synchrotron Light Source II (NSLS‐II) is presented. Designed for hard X‐ray photon‐in/photon‐out spectroscopy, the spectrometer achieves a resolution in the 0.5–2 eV range, depending on the element and/or emission line, providing detailed insights into the local electronic and geometric structure of materials. It serves a diverse user community, including fields such as physical, chemical, biological, environmental and materials sciences. This article details the mechanical design, alignment procedures and data‐acquisition scheme of the spectrometer, with a particular focus on the continuous asynchronous data‐acquisition approach that significantly enhances experimental efficiency. [ABSTRACT FROM AUTHOR] |
قاعدة البيانات: |
Academic Search Index |