CENTRAL CORNEAL THICKNESS MEASURED BY THE OCULYZER, BIOGRAPH, AND ULTRASOUND PACHYMETRY. |
Alternate Title: |
CENTRALNA DEBLJINA ROŽNJAČE MERENA OCULYZEROM, BIOGRAFOM I ULTRAZVUCNIM PAHIMETROM. |
المؤلفون: |
Zlatanović, Marko1, Živković, Maja1,2, Hristov, Aleksandra3 alexandramiljkovic@gmail.com, Stojković, Vesna3, Novak, Saša1, Zlatanović, Nevena2, Brzaković, Mladen3 |
المصدر: |
Acta Medica Medianae. Jun2019, Vol. 58 Issue 2, p33-37. 5p. 1 Chart. |
مصطلحات موضوعية: |
*ULTRASONIC imaging, *INTRAOCULAR pressure, *PATIENT monitoring, *STANDARD deviations, *DYSTROPHY |
Abstract (English): |
Central corneal thickness (CCT) is an important parameter in selecting patients before any corneal refractive procedures and in postoperative follow-up, in monitoring the patients with corneal disorders such as keratoconus and Fuchs' dystrophy, as well as in adjusting the intraocular pressure according to central corneal thickness. The aim of the paper was to compare the results of CCT measurements obtained by different measurement instruments: Oculyzer, BioGraph and ultrasound pachymetry. CCT was measured in 64 eyes of 32 patients. The measurements were performed in a specialized Eye Hospital "Maja Clinic" in Niš, using the following instruments: WaveLight Allegro Oculyzer, WaveLight Allergo Biograph, ultrasound pachymeter DGH Pachette 3. The mean values of central corneal thickness obtained from different devices were compared. The mean values of CCT (± standard deviation) obtained with Oculyzer were 552.94 µm ± 22.88 µm, 556.56 ± 25.32 µm obtained with BioGraph, and 559.46 ± 26.0 µm measured by ultrasound pachymeter. There were no statistically significant differences among different measurement devices. The CCT measurements with the Oculyzer, BioGraph, and ultrasound pachymetry do not show statistically significant differences, so the results of CCT measurements obtained by using any of these devices may be considered valid. [ABSTRACT FROM AUTHOR] |
Abstract (Bosnian): |
Centralna debljina rožnjače (CCT) je parametar bitan u selekciji bolesnika kod svih kornealnih refraktivnih procedura i u postoperativnom praćenju, u praćenju pacijanata sa bolestima rožnjače kao što su keratokonus i Fuksova distrofija, kao i u korekciji izmerenog intraokularnog pritiska. Cilj rada bio je uporediti rezultate merenja CCT dobijene različitim instrumentima merenja: oculyzerom, biografom i ultrazvučnim pahimetrom. Merena je CCT kod 32 bolesnika kod 64 očiju. Merenje je vršeno u specijalnoj očnoj bolnici "Klinika Maja" u Nišu na instrumentima: WaveLight Allegro Oculyzer, WaveLight Allergo Biograph, ultrazvučni pahimetar DGH Pachette 3. Upoređivane su srednje vrednosti CCT merene različitim instrumentima. Dobijene srednje vrednosti CCT (± standardna devijacija) iznosile su 552,94 µm ± 22,88 µm mereno Oculyzerom, 556,56 ± 25,32 µm mereno Biografom i 559,46 ± 26,0 µm mereno ultrazvučnim pahimetrom, pri čemu dobijene razlike među različitim aparatima nisu statistički značajne. Rezultati merenja CCT oculyzerom, biografom i ultrazvučnim pahimetrom ne pokazuju statistički značajne razlike, te se rezultati CCT dobijeni na bilo kom od ovih instrumenata mogu smatrati validnim. [ABSTRACT FROM AUTHOR] |