Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study
العنوان: | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
---|---|
المؤلفون: | Werner Baulig, Samira Akbas, Philipp K. Schütt, Wolfgang Keul, Marija Jovic, Pascal Berdat, Stefanie von Felten, Klaus Steigmiller, Michael Thomas Ganter, Oliver M. Theusinger |
المصدر: | BMC Anesthesiology, Vol 21, Iss 1, Pp 1-12 (2021) |
بيانات النشر: | BMC |
سنة النشر: | 2021 |
المجموعة: | Directory of Open Access Journals: DOAJ Articles |
مصطلحات موضوعية: | Resonance Sonorheometry, Viscoelastic testing, Quantra, ROTEM sigma, cardiac surgery, Anesthesiology, RD78.3-87.3 |
الوصف: | Background Measures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA). Methods In thirty-eight patients scheduled for elective cardiac surgery between December 2018 and October 2019, blood samples were taken after induction of anesthesia (sample 1) and after heparin neutralization (sample 2) and measured on Quantra (QPlus® Cartridge) and ROTEM sigma (ROTEM® sigma complete + hep Cartridge). Clot times and clot stiffness values were recorded. Clot stiffness values of ROTEM amplitudes (A in mm) were converted to shear modulus (G) in hectoPascal (hPa): G (hPa) = (5 x A)/(100-A). Additionally, time-to-results was recorded. Spearman rank test correlation and Bland Altman analysis were performed. Results Clot stiffness parameters of the Quantra correlated strongly with corresponding measurements of the ROTEM with r = 0.93 and 0.94 for EXTEM A10 vs CS and r = 0.94 and 0.96 for FIBTEM A10 vs FCS for sample 1 and 2, respectively. Quantra clot time correlated strongly with ROTEM INTEM CT with r = 0.71 for sample 1 and r = 0.75 for sample 2. However, Bland Altman analysis showed no agreement in all compared assays of both methods. The median time to delivery of first and complete results was significantly shorter for Quantra (412 and 658 s) compared to ROTEM sigma (839 and 1290 s). Conclusions The Quantra showed a strong correlation with the ROTEM sigma for determining clot times and clot stiffness and the parameters assess similar aspects of clot development. However, these parameters are not directly interchangeable and implicate that separate cut-off values need to be established for users of the Quantra device. Word count: 278. Trial registration The study was retrospectively registered with ClinicalTrials.gov (ID: NCT04210830 ) at December 20th 2019. |
نوع الوثيقة: | article in journal/newspaper |
اللغة: | English |
تدمد: | 1471-2253 |
Relation: | https://doi.org/10.1186/s12871-021-01469-5; https://doaj.org/toc/1471-2253; https://doaj.org/article/9f3e20f7e63a47a58d47103e92223bd1 |
DOI: | 10.1186/s12871-021-01469-5 |
الاتاحة: | https://doi.org/10.1186/s12871-021-01469-5 https://doaj.org/article/9f3e20f7e63a47a58d47103e92223bd1 |
رقم الانضمام: | edsbas.1116842A |
قاعدة البيانات: | BASE |
ResultId |
1 |
---|---|
Header |
edsbas BASE edsbas.1116842A 919 3 Academic Journal academicJournal 918.9462890625 |
PLink |
https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&scope=site&db=edsbas&AN=edsbas.1116842A&custid=s6537998&authtype=sso |
FullText |
Array
(
[Availability] => 0
)
Array ( [0] => Array ( [Url] => https://doi.org/10.1186/s12871-021-01469-5# [Name] => EDS - BASE [Category] => fullText [Text] => View record in BASE [MouseOverText] => View record in BASE ) [1] => Array ( [Url] => https://resolver.ebscohost.com/openurl?custid=s6537998&groupid=main&authtype=ip,guest&sid=EBSCO:edsbas&genre=article&issn=14712253&ISBN=&volume=&issue=&date=20210101&spage=&pages=&title=BMC Anesthesiology&atitle=Comparison%20of%20the%20resonance%20sonorheometry%20based%20Quantra%C2%AE%20system%20with%20rotational%20thromboelastometry%20ROTEM%C2%AE%20sigma%20in%20cardiac%20surgery%20%E2%80%93%20a%20prospective%20observational%20study&id=DOI:10.1186/s12871-021-01469-5 [Name] => Full Text Finder (s6537998api) [Category] => fullText [Text] => Full Text Finder [Icon] => https://imageserver.ebscohost.com/branding/images/FTF.gif [MouseOverText] => Full Text Finder ) ) |
Items |
Array
(
[Name] => Title
[Label] => Title
[Group] => Ti
[Data] => Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study
)
Array ( [Name] => Author [Label] => Authors [Group] => Au [Data] => <searchLink fieldCode="AR" term="%22Werner+Baulig%22">Werner Baulig</searchLink><br /><searchLink fieldCode="AR" term="%22Samira+Akbas%22">Samira Akbas</searchLink><br /><searchLink fieldCode="AR" term="%22Philipp+K%2E+Schütt%22">Philipp K. Schütt</searchLink><br /><searchLink fieldCode="AR" term="%22Wolfgang+Keul%22">Wolfgang Keul</searchLink><br /><searchLink fieldCode="AR" term="%22Marija+Jovic%22">Marija Jovic</searchLink><br /><searchLink fieldCode="AR" term="%22Pascal+Berdat%22">Pascal Berdat</searchLink><br /><searchLink fieldCode="AR" term="%22Stefanie+von+Felten%22">Stefanie von Felten</searchLink><br /><searchLink fieldCode="AR" term="%22Klaus+Steigmiller%22">Klaus Steigmiller</searchLink><br /><searchLink fieldCode="AR" term="%22Michael+Thomas+Ganter%22">Michael Thomas Ganter</searchLink><br /><searchLink fieldCode="AR" term="%22Oliver+M%2E+Theusinger%22">Oliver M. Theusinger</searchLink> ) Array ( [Name] => TitleSource [Label] => Source [Group] => Src [Data] => BMC Anesthesiology, Vol 21, Iss 1, Pp 1-12 (2021) ) Array ( [Name] => Publisher [Label] => Publisher Information [Group] => PubInfo [Data] => BMC ) Array ( [Name] => DatePubCY [Label] => Publication Year [Group] => Date [Data] => 2021 ) Array ( [Name] => Subset [Label] => Collection [Group] => HoldingsInfo [Data] => Directory of Open Access Journals: DOAJ Articles ) Array ( [Name] => Subject [Label] => Subject Terms [Group] => Su [Data] => <searchLink fieldCode="DE" term="%22Resonance+Sonorheometry%22">Resonance Sonorheometry</searchLink><br /><searchLink fieldCode="DE" term="%22Viscoelastic+testing%22">Viscoelastic testing</searchLink><br /><searchLink fieldCode="DE" term="%22Quantra%22">Quantra</searchLink><br /><searchLink fieldCode="DE" term="%22ROTEM+sigma%22">ROTEM sigma</searchLink><br /><searchLink fieldCode="DE" term="%22cardiac+surgery%22">cardiac surgery</searchLink><br /><searchLink fieldCode="DE" term="%22Anesthesiology%22">Anesthesiology</searchLink><br /><searchLink fieldCode="DE" term="%22RD78%2E3-87%2E3%22">RD78.3-87.3</searchLink> ) Array ( [Name] => Abstract [Label] => Description [Group] => Ab [Data] => Background Measures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA). Methods In thirty-eight patients scheduled for elective cardiac surgery between December 2018 and October 2019, blood samples were taken after induction of anesthesia (sample 1) and after heparin neutralization (sample 2) and measured on Quantra (QPlus® Cartridge) and ROTEM sigma (ROTEM® sigma complete + hep Cartridge). Clot times and clot stiffness values were recorded. Clot stiffness values of ROTEM amplitudes (A in mm) were converted to shear modulus (G) in hectoPascal (hPa): G (hPa) = (5 x A)/(100-A). Additionally, time-to-results was recorded. Spearman rank test correlation and Bland Altman analysis were performed. Results Clot stiffness parameters of the Quantra correlated strongly with corresponding measurements of the ROTEM with r = 0.93 and 0.94 for EXTEM A10 vs CS and r = 0.94 and 0.96 for FIBTEM A10 vs FCS for sample 1 and 2, respectively. Quantra clot time correlated strongly with ROTEM INTEM CT with r = 0.71 for sample 1 and r = 0.75 for sample 2. However, Bland Altman analysis showed no agreement in all compared assays of both methods. The median time to delivery of first and complete results was significantly shorter for Quantra (412 and 658 s) compared to ROTEM sigma (839 and 1290 s). Conclusions The Quantra showed a strong correlation with the ROTEM sigma for determining clot times and clot stiffness and the parameters assess similar aspects of clot development. However, these parameters are not directly interchangeable and implicate that separate cut-off values need to be established for users of the Quantra device. Word count: 278. Trial registration The study was retrospectively registered with ClinicalTrials.gov (ID: NCT04210830 ) at December 20th 2019. ) Array ( [Name] => TypeDocument [Label] => Document Type [Group] => TypDoc [Data] => article in journal/newspaper ) Array ( [Name] => Language [Label] => Language [Group] => Lang [Data] => English ) Array ( [Name] => ISSN [Label] => ISSN [Group] => ISSN [Data] => 1471-2253 ) Array ( [Name] => NoteTitleSource [Label] => Relation [Group] => SrcInfo [Data] => https://doi.org/10.1186/s12871-021-01469-5; https://doaj.org/toc/1471-2253; https://doaj.org/article/9f3e20f7e63a47a58d47103e92223bd1 ) Array ( [Name] => DOI [Label] => DOI [Group] => ID [Data] => 10.1186/s12871-021-01469-5 ) Array ( [Name] => URL [Label] => Availability [Group] => URL [Data] => https://doi.org/10.1186/s12871-021-01469-5<br />https://doaj.org/article/9f3e20f7e63a47a58d47103e92223bd1 ) Array ( [Name] => AN [Label] => Accession Number [Group] => ID [Data] => edsbas.1116842A ) |
RecordInfo |
Array
(
[BibEntity] => Array
(
[Identifiers] => Array
(
[0] => Array
(
[Type] => doi
[Value] => 10.1186/s12871-021-01469-5
)
)
[Languages] => Array
(
[0] => Array
(
[Text] => English
)
)
[Subjects] => Array
(
[0] => Array
(
[SubjectFull] => Resonance Sonorheometry
[Type] => general
)
[1] => Array
(
[SubjectFull] => Viscoelastic testing
[Type] => general
)
[2] => Array
(
[SubjectFull] => Quantra
[Type] => general
)
[3] => Array
(
[SubjectFull] => ROTEM sigma
[Type] => general
)
[4] => Array
(
[SubjectFull] => cardiac surgery
[Type] => general
)
[5] => Array
(
[SubjectFull] => Anesthesiology
[Type] => general
)
[6] => Array
(
[SubjectFull] => RD78.3-87.3
[Type] => general
)
)
[Titles] => Array
(
[0] => Array
(
[TitleFull] => Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study
[Type] => main
)
)
)
[BibRelationships] => Array
(
[HasContributorRelationships] => Array
(
[0] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Werner Baulig
)
)
)
[1] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Samira Akbas
)
)
)
[2] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Philipp K. Schütt
)
)
)
[3] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Wolfgang Keul
)
)
)
[4] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Marija Jovic
)
)
)
[5] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Pascal Berdat
)
)
)
[6] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Stefanie von Felten
)
)
)
[7] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Klaus Steigmiller
)
)
)
[8] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Michael Thomas Ganter
)
)
)
[9] => Array
(
[PersonEntity] => Array
(
[Name] => Array
(
[NameFull] => Oliver M. Theusinger
)
)
)
)
[IsPartOfRelationships] => Array
(
[0] => Array
(
[BibEntity] => Array
(
[Dates] => Array
(
[0] => Array
(
[D] => 01
[M] => 01
[Type] => published
[Y] => 2021
)
)
[Identifiers] => Array
(
[0] => Array
(
[Type] => issn-print
[Value] => 14712253
)
[1] => Array
(
[Type] => issn-locals
[Value] => edsbas
)
[2] => Array
(
[Type] => issn-locals
[Value] => edsbas.oa
)
)
[Titles] => Array
(
[0] => Array
(
[TitleFull] => BMC Anesthesiology
[Type] => main
)
)
)
)
)
)
)
|
IllustrationInfo |