Particle Size Measurement Using Dynamic Light Scattering at Ultra-Low Concentration Accounting for Particle Number Fluctuations.

التفاصيل البيبلوغرافية
العنوان: Particle Size Measurement Using Dynamic Light Scattering at Ultra-Low Concentration Accounting for Particle Number Fluctuations.
المؤلفون: Wang, Mengjie, Shen, Jin, Thomas, John C., Mu, Tongtong, Liu, Wei, Wang, Yajing, Pan, Jinfeng, Wang, Qin, Liu, Kaishi
المصدر: Materials (1996-1944); Oct2021, Vol. 14 Issue 19, p5683, 1p
مصطلحات موضوعية: PARTICLE size determination, LIGHT scattering, PARTICLE size distribution, KERNEL functions, LIGHT intensity
مستخلص: Dynamic light scattering (DLS) is a popular method of particle size measurement, but at ultra-low particle concentrations, the occurrence of number concentration fluctuations limits the use of the technique. Number fluctuations add a non-Gaussian term to the scattered light intensity autocorrelation function (ACF). This leads to an inaccurate particle size distribution (PSD) being recovered if the normal DLS analysis model is used. We propose two methods for inverting the DLS data and recovering the PSDs when number fluctuations are apparent. One is to directly establish the relationship between the non-Gaussian ACF and the PSD by the kernel function reconstruction (KFR) method while including the non-Gaussian term to recover the PSD. The other is to remove the effect of the non-Gaussian term in the ACF by the baseline reset (BR) method. By including the number fluctuation term, the ideal recovered PSD can be obtained from the simulated data, but this will not happen in the experimental measurement data. This is because the measured intensity ACF contains more noise than the simulated ACF at ultra-low concentration. In particular, the baseline noise at the tail of long delay time of ACF overwhelms the number fluctuation term, making it difficult to recover reliable PSD data. Resetting the baseline can effectively remove the digital fluctuation term in ACF, which is also a feasible method to improve PSD recovery under ultra-low concentration. However, increasing noise at ultra-low concentrations can lead to errors in determining an effective baseline. This greatly reduces the accuracy of inversion results. Results from simulated and measured ACF data show that, for both methods, noise on the ACF limits reliable PSD recovery. [ABSTRACT FROM AUTHOR]
Copyright of Materials (1996-1944) is the property of MDPI and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:19961944
DOI:10.3390/ma14195683